Description
Sensing element
Thermocouple type K, J or T, single or duplex with tolerance class according to IEC 60584 (EN 584).
Temperature range
0 to 600 °C
EXTERNAL SHEATH
AISI 316L seamless tube – wall thickness 1 mm
Diameter: 9 mm standard, other diameters available on request
Insertion length: standard 160, 250 and 400 mm,any other length on request.
Hint: Insertion length
For parallel thread (BSP) the length L is measured WITH the thread.
With conical threads (BSPT and NPT) the length L is measured WITHOUT the thread.
Process connection
“Standard ½”” BSP-M
Other sizes or threads available on request.”
EXTENSION OR LAGGING
Standard 145 mm, other lengths available on request
CONNECTION HEAD
Standard DIN type B head with M20 cable gland (other connection heads available on request)
Applications
General purpose measurement probe – all industry domains.
Options
4-20 mA or 0/10V transmitter (insulated, non insulated, HART, etc.) Can be built in only if connection head or ambient temperature does not exceed 75 °C - Ex Design
Characteristics
Mechanical strength of probes with threaded connector
Flow | Insertion length | For stainless steel at | |||||
---|---|---|---|---|---|---|---|
Water | Air | 50°C | 300°C | 400°C | |||
3m/s | 25m/s | 160 mm | 50 kg/cm² | 40 kg/cm² | 36 kg/cm² | ||
250 mm | 40 kg/cm² | 40 kg/cm² | 36 kg/cm² | ||||
450 mm | 18 kg/cm² | 18 kg/cm² | 18 kg/cm² |